Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0743-2025

Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837 Recall

Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837

FDA recall Class II

Ongoing

FDA lists this Class II recall for Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0743-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 22, 2024
FDA report date
Jan 1, 2025
Quantity in commerce
171 orders
Distribution
See key details
FDA cause
Software design (manufacturing process)

What is being recalled?

Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Only 3M Clarity Precision Grip Attachment Dual Arch Attachments are potentially affected. Part numbers/UDI-DI: 359-830/ 00652221307286; 359-831/ 00652221307293; 359-832/00652221307309; 359-833/ 00652221307316; 359-834/ 00652221307361; 359-835/ 00652221307378; 359-836/ 00652221307385; 359-837/ 00652221307392

View the original FDA recall notice

Why was this product recalled?

Attachments may not mate with the tooth as intended in the digital treatment design.

What actions did 3M Company - Health Care Business take?

On November 22, 2024, the firm notified customers via URGENT Product Recall letters. Customers were informed that select grip attachments were impacted by a software nonconformance, limited to the lower arch of specific dual-arch Clarity Precision Grip Attachment orders. The firm advises that customers not use products from the identified Case IDs, and that they should dispose of affected products. Customers should complete and return the included customer product destruction form. The firm will provide replacement product. If you have additional questions or need to report an adverse event, contact the Clinician Support team (800-423-4588, option 2) or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA classification recorded

  3. FDA notice posted

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.