Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0744-2025

TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis. Recall

TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.

FDA recall Class II

Ongoing

FDA lists this Class II recall for TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0744-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 2, 2024
FDA report date
Jan 1, 2025
Quantity in commerce
~87 devices (U.S.)
Distribution
See key details
FDA cause
Process control

What is being recalled?

TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.

Affected model & UDI checker

Search the exact identifiers published for Z-0744-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1590.15.010TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.Check source noticeNot specified
1590.15.020TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.Check source noticeNot specified
Not specifiedTEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.Check source notice08033390128865
Not specifiedTEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.Check source notice08033390128872
FDA code information

Product/Catalog Numbers: 1590.15.010 TEMA ELBOW - AXLE SMALL UDI-DI code: 08033390128865 Lot/Serial Numbers: 23031752300054 23177372300230 23173852400087 Product/Catalog Number: 1590.15.020 TEMA ELBOW - AXLE LARGE UDI-DI code: 08033390128872 Lot/Serial Numbers: 22276962300054 23325482400047 24006192400193

View the original FDA recall notice

Why was this product recalled?

Due to a potential difference in mechanical behavior between the implanted device and tool(s) used during revision surgery.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Limacorporate S.p.A. take?

On 10/02/2024, the firm notified U.S. surgeons of the recall and the two U.S. distributors were informed via email on 10/11/2024. An updated/follow-up Recall Notification and Acknowledgement form is pending. The update/follow-up Recall Notification is informing surgeons and distributors that the firm has received a total of two similar complaints, regarding issues related to difficulty in disassembly of the axle component by using the screwdriver as per Surgical Technique. According to the internal investigation, the main concurring root causes of the problem seem to be the incomplete shot peening on the taper and the missing identification of a critical dimension (thread length). The Recall Notification informs surgeons to use other instruments available in the Operating Room to successfully complete the surgery or, in the worst case, several parts/the entire implant (both the humeral and the ulnar components) should be removed. Surgeons are also asked to Fill out, sign, and send the following Acknowledgement and Receipt Form via email as confirmation that You have read and acknowledged the content of this document. For questions or assistance, contact: Contact of the Importer: Kenneth Newman - Senior Regulatory Affairs Specialist (LimaCorporate now Enovis) /Recall Coordinator (Enovis) E-mail: [email protected] Or to the Manufacturer contact point: Federica Malvaso Post-Market Surveillance Lead (LimaCorporate now Enovis) E-mail: [email protected] Distributors are instructed to: Check your stock to locate and quarantine the affected devices. Devices must be sent back to: Lima U.S.A., Inc. | 2001 NE Green Oaks Blvd. Ste.100 | Arlington, Texas 76006 | USA within 15 days together with a signed hard copy of the following Acknowledgement and Receipt Form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology