Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0767-2025

Magnesium Flex reagent cartridge, 360 tests Recall

Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Magnesium Flex reagent cartridge, 360 tests. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0767-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 12, 2024
FDA report date
Jan 1, 2025
Quantity in commerce
13,115 units (11,520 US, 1,595 ROW)
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;

Affected model & UDI checker

Search the exact identifiers published for Z-0767-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
K3057Magnesium Flex reagent cartridge, 360 testsCheck source noticeNot specified
Not specifiedMagnesium Flex reagent cartridge, 360 testsCheck source notice00842768015694
FDA code information

Siemens Material Number (SMN): 10445158; Catalog Number: K3057; UDI-DI: 00842768015694; Lot numbers: 24057BB, 24064BC.

View the original FDA recall notice

Why was this product recalled?

Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Healthcare Diagnostics, Inc. take?

On November 12, 2024 URGENT MEDICAL DEVICE CORRECTION letters were sent to consignees. Customer Actions Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Discontinue use of and discard the lots listed above. Review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within thirty (30) days. Please retain this letter with your laboratory records and forward this letter to those who may have received the product. Resolution Siemens Healthineers has introduced additional manufacturing process controls that have resolved this issue. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology