Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0804-2025

Smaxel Fractional CO2 Laser, a Class IV medical laser system. Recall

Smaxel Fractional CO2 Laser, a Class IV medical laser system.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Smaxel Fractional CO2 Laser, a Class IV medical laser system.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0804-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 16, 2024
FDA report date
Jan 15, 2025
Quantity in commerce
23
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

Smaxel Fractional CO2 Laser, a Class IV medical laser system.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

unknown

View the original FDA recall notice

Why was this product recalled?

The Smaxel Fractional CO2 Laser does not include a Certification label, and adequate instructions for safe operation of the device.

What actions did IDS LTD take?

IDS LTD recommend that users are advised to use it as close to the main body as possible during the procedure. The firm will without charge remedy the defect or bring the product into compliance with each applicable Federal standard. On or about 1/22/2025, the firm mailed a "Notification Letter" to affected customers/users informing them that the SAXEL Fractional CO2 Laser System does not have affixed a certification label and that the firm will be implementing a design change by revising the certification label and will provide the revised certification label free of charge following the design change.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.