Smaxel Fractional CO2 Laser, a Class IV medical laser system. Recall
Smaxel Fractional CO2 Laser, a Class IV medical laser system.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Smaxel Fractional CO2 Laser, a Class IV medical laser system.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0804-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 16, 2024
- FDA report date
- Jan 15, 2025
- Quantity in commerce
- 23
- Distribution
- See key details
- FDA cause
- Radiation Control for Health and Safety Act
What is being recalled?
Smaxel Fractional CO2 Laser, a Class IV medical laser system.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
unknown
Why was this product recalled?
The Smaxel Fractional CO2 Laser does not include a Certification label, and adequate instructions for safe operation of the device.
What actions did IDS LTD take?
IDS LTD recommend that users are advised to use it as close to the main body as possible during the procedure. The firm will without charge remedy the defect or bring the product into compliance with each applicable Federal standard. On or about 1/22/2025, the firm mailed a "Notification Letter" to affected customers/users informing them that the SAXEL Fractional CO2 Laser System does not have affixed a certification label and that the firm will be implementing a design change by revising the certification label and will provide the revised certification label free of charge following the design change.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0804-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology