Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile. Recall
Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0808-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 12, 2024
- FDA report date
- Jan 15, 2025
- Quantity in commerce
- 479 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.
Affected model & UDI checker
Search the exact identifiers published for Z-0808-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| CAT1785 | Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile. | Check source notice | Not specified |
| CAT500 | Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile. | Check source notice | Not specified |
| Not specified | Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile. | Check source notice | 08886479300269 |
FDA code information
Lot #10235, exp. 2026-05-01; UDI (01)08886479300269(17)260501(10)10235
Why was this product recalled?
Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.
What actions did Aerin Medical, Inc. take?
The recalling firm issued emails on 11/12/2024. It was necessary to notify several consignees by mail with an "URGENT: MEDICAL DEVICE FIELD NOTICE" letter dated 11/20/2024. The email described the product purchased by the consignee and the reason for recall. The consignee is informed the programmed treatment time will be shorter than the recommended 30 second treatment time for VivAer and the use of the device may result in insufficient treatment. Under the heading "Action to be Taken," it says if a defective device is identified or the consignee wishes to return unused devices within this production lot, they are to return them to the recalling firm, and return instructions are provided. Regarding the mailed "URGENT: MEDICAL DEVICE FIELD NOTICE" letters dated 11/20/2024, they informed the consignee that some of the units in the production lot were programmed incorrectly and, in these cases, the programmed treatment time will be 12 seconds of radiofrequency energy delivery with no cooling time, not the 30 second treatment time for VivAer that includes 18 seconds of energy delivery and 12 seconds of cooling time. The letter also contains instructions to share this notification with all device users within their facility and network to ensure they are aware of this issue. This notification should also be shared with any organization where the potentially affected devices have been transferred. If you need any further information or support concerning this notification, please contact us at [email protected] or call 512-221-9956.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0808-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology