Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data. Recall
Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0810-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 25, 2024
- FDA report date
- Jan 15, 2025
- Quantity in commerce
- 434
- Distribution
- See key details
- FDA cause
- Software change control
What is being recalled?
Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.
Affected model & UDI checker
Search the exact identifiers published for Z-0810-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 96280 | Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data. | Check source notice | Not specified |
| Not specified | Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data. | Check source notice | 10841522107177 |
| Not specified | Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data. | Check source notice | 10841522100246 |
FDA code information
Software Versions: 1.4.0 and 1.4.1. UDI-DI: 10841522107177, 10841522100246
Why was this product recalled?
Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
What actions did Spacelabs Healthcare, Inc. take?
On 11/25/2024, correction notices were emailed to customers who were asked to do the following: Firm will contact you at the earliest possible date to schedule a convenient time to update affected devices to software version 1.4.2 for XTR systems which addresses this defect. You can also contact GTS to have an FSE paged out for discussing upgrade schedule: 1-800-522-7025 and select 2 for Technical Support. Complete and submit the customer acknowledgement form: https://spacelabshealthcare.com/support/recall-reply-form/recall-reply-form-xtr-15-nov-2024/ For additional information or technical assistance, or to report an adverse event, contact: Technical Support: 1-800-522-7025 and select 2
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0810-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology