FDA RECALL RECORD · Z-0811-2025

GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB Recall

GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0811-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 25, 2024
FDA report date
Jan 15, 2025
Quantity in commerce
53 units
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB

Affected model & UDI checker

Search the exact identifiers published for Z-0811-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedGE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MBCheck source notice00840682115278
FDA code information

1) 5555000-3: UDI/DI 00840682115278, Serial Numbers: 260665XR220, 478757OPT, 512SDERLM1, 904244MP5, 912466AMX2, 954WRMCDP2; UDI/DI none, Serial Numbers: 10523MOB01, 201418DIG1, 210297NC200M1, 229276M200, 240566P1, 316858AMX200P1, 401OPTI3, 410368WP1, 443643WP1, 503251XR200, 512SDMCP1, 518243OPT2, 603356OPTIMA, 609267OPT, 609267OPT3, 661949DP1, 661949DP2, 718630OPTIMAX, 760379DP, 801299SB1, 804417OPTIMA, 812237P1, 813745OPT1, 856355OPT, 904244MP1, 904244MP2, 904276DP200, 918477OPT1, 956365OPT1, 956632OPT2, 956632OPT4, XR1037005WK4; 2) 5555000-4: UDI/DI 00840682115278, Serial Numbers: 405717P200, 408259AP8, 518583OPT1, 773792XR200, 7982XR0277, 801230XRA, 904399MP19, 989907DP1; UDI/DI none, Serial Numbers: 1553XR0081, 316962240P1, 361572XR220C, 812265M200, A51043204, FEB520100; 3) A0659MB: UDI/DI 00840682115278, Serial Numbers: XR2262CO.

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.

What actions did GE Medical Systems, LLC take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/25/2024 via letter issued using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken by the user: "1. Perform a check of your device prior to each patient exam to look for the following behaviors. If any are observed, DO NOT use the device and call GE HealthCare Service Immediately. . Column vertical movement requires increased effort. . Column has change in resistance (i.e. movement is not smooth) during vertical movement. . Column makes abnormal noises during vertical movement. 2. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. 3. Please retain this document for your records. 4. Please complete and return the attached acknowledgement form to [email protected]" For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. On 02/11/2025, the firm send an updated notice with a corrected GTIN for Optima XR200amx.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology