GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided. Recall
GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided.
FDA recall Class II
Ongoing
FDA lists this Class II recall for GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0813-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 25, 2024
- FDA report date
- Jan 15, 2025
- Quantity in commerce
- 120 units
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided.
Affected model & UDI checker
Search the exact identifiers published for Z-0813-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided. | Check source notice | 840682122351 |
FDA code information
1) 5555000-27: UDI/DI 840682122351, Serial Numbers: 228867XR240, 252847PP241, 281274OPT1, 281338CL240, 281VHVXR240, 305682DIG1, 305682DIG3, 305KEN240B, 401OPTI2, 405949OPT1, 408559XRDR7, 413447OPT1, 419251OPT, 419690DP1, 504842XR242, 602344240A, 602344MARY240, 713440DP2, 713440DP3, 720854DF1, 732222DP3, 763520XR10, 763520XRA6, 806725OPT, 865305DP1, 941276XRA, 970495OPTI1, 973926OPT1, BOLDWPACPORT02, GLEDWPACPORT02, HMHWT240P5, LIHOPTIMA240B, OCHSNERWB240, SAMC240A; 2) 5555000-31: UDI/DI 840682122351, Serial Numbers: 00158MOB17, DF2401800221WK, U416249MOB240; 3) 5555000-33: UDI/DI 840682122351, Serial Numbers: 082407100689, 228867XR240C; 4) 5555000-41: UDI/DI 840682122351, Serial Numbers: 00130MOB16, 5973XR0225, DEF11034, RS4025RX05, SA1058MM50; 5) 5555000-47: UDI/DI 840682122351, Serial Numbers: 206223OPTIMA1, 206248OPT2, 206341OPT2402, 212838OPTIMA1, 219326AMX2, 256HHPORT5, 256HHPORT6, 260824XR240, 270793OPT, 270793OPT1, 304256OPTIMA240, 307324XR240, 318626OPX240FP, 336716WFP2, 336716WFP3, 404616OPT2, 405657OPT, 406345OPT1, 406653XR240, 406653XRA1, 414266XR243, 423837X240, 425228OPT240, 435713OPX240, 435789XR240, 502897M241, 504842OPTFP1, 504894XR244, 513952OPT1, 530888OPT2, 541789M3, 551996DIG1, 706527OPT2, 712662OPT240, 713440DP4, 718918XR240B, 785623OPT2, 786308XR240, 801465XR240, 805652M2, 816943XR240A, 828369OPT1, 850864FWDH240, 865977DP2, 901595XRA1, 914787XR2401, 916781OPT4, 951788XR240P2, 954723XR240B, 972596OPT3, 973877OPT1, AMX240GEN2SYSTEM17, AMX240GEN2SYSTEM9, HMHXR240P7, KAFB228OPTIMA2, LNMCDP1, LVCCER240B, MBMCPORTD; 6) A0700DL: UDI/DI 840682122351, Serial Numbers: 082407250223; 7) A1700DA: UDI/DI 840682122351, Serial Numbers: 210METM2, 256543OPT, 337261OPT, 360744OPT2402, 425899OPT, 425SNORIDGE, 509946AMX1, 573814DP8, 574231OPT2, 702852OPT, 850864FWDH2401, HMHXR240P1, OMCTOPT1; 8) To be provided: UDI/DI To be provided, Serial Numbers: 256543OPT1, 405657FP1417, 513952OPT2, RU1018RX04.
Why was this product recalled?
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
What actions did GE Medical Systems, LLC take?
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/25/2024 via letter issued using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken by the user: "1. Perform a check of your device prior to each patient exam to look for the following behaviors. If any are observed, DO NOT use the device and call GE HealthCare Service Immediately. . Column vertical movement requires increased effort. . Column has change in resistance (i.e. movement is not smooth) during vertical movement. . Column makes abnormal noises during vertical movement. 2. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. 3. Please retain this document for your records. 4. Please complete and return the attached acknowledgement form to [email protected]" For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. On 02/11/2025, the firm send an updated notice with a corrected GTIN for Optima XR200amx.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0813-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology