AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port. Recall
AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port. Model/Catalog Number: E6-1222
FDA recall Class II
Completed
FDA lists this Class II recall for AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0819-2025
- Recall classification
- Class II
- Enforcement status
- Completed
- Initiation date
- Nov 26, 2024
- FDA report date
- Jan 15, 2025
- Quantity in commerce
- 100
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port. Model/Catalog Number: E6-1222
Affected model & UDI checker
Search the exact identifiers published for Z-0819-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| E6-1222 | AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port. | Check source notice | Not specified |
| Not specified | AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port. | Check source notice | 00842071107086 |
FDA code information
Lot Code: Model No: E6-1222 UDI-DI: (01)00842071107086(17)270401(10)240424-628 Lot number: 240424-628 Exp. Date: 2027-04-01
Why was this product recalled?
The extension sets subject to this recall have ENFit connections, although they should have non- ENFit connections. ENFit connectors will only fit with other ENFit-compatible devices, and will not fit with non-ENFit connectors.
What actions did Applied Medical Technology Inc take?
An "URGENT: MEDICAL DEVICE RECALL" notification letter dated 11/26/24 was sent to customers. ACTIONS TO BE TAKEN: CUSTOMER OR USER: Applied Medical Technology, Inc. requests that customers IMMEDIATELY TAKE THE FOLLOWING ACTIONS: " Examine your inventory and isolate/quarantine pending return of the product subject to this recall. " Do not use any device subject to this recall. " Complete and return to Applied Medical Technology, Inc. the enclosed RECALL NOTIFICATION ACKNOWLEDGMENT via fax or send by mail. " Please contact Applied Medical Technology, Inc. either by phone at 440-717-4000, or via email at [email protected], to make arrangements for the return of any devices related to this recall. Include the product, all of its components, and any available packaging and labeling associated with the device. An RMA number will only be issued for pouches with lot number 240424-628. Please contact us with any questions or concerns: Public Recall Contact: Customer Service Address: Applied Medical Technology, Inc. Attn: Customer Service 8006 Katherine Blvd Brecksville, OH 44141 USA Telephone: 440-717-4000 Fax: 440-717-4200 Email: [email protected] General hours of operation are 8:30am to 5:00pm. All times in Eastern Standard Time Zone.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0819-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026