Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges. Recall
Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0836-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 10, 2024
- FDA report date
- Jan 22, 2025
- Quantity in commerce
- 8 units
- Distribution
- See key details
- FDA cause
- Environmental control
What is being recalled?
Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges.
Affected model & UDI checker
Search the exact identifiers published for Z-0836-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| P0010 | Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges. | Check source notice | Not specified |
| Not specified | Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges. | Check source notice | 05415219000119 |
FDA code information
UDI/DI 05415219000119, Serial Numbers: 00002706, 00002820, 00003103, 00003262, 00003419, 00003434, 00003444, and 00003445.
Why was this product recalled?
False-positive MSI-H results generated by the Idylla MSI Test.
What actions did Biocartis Nv take?
Biocartis issued an URGENT - Medical Device Correction notice to its consignees on 12/10/2024 via email. The notice provided the description of the problem, its impact, and requested the following actions be taken: 1. Until your Instrument is serviced, please move forward with the option that applies to your situation: o In case you have another Idylla Instrument available that is not affected, use that Instrument to run the Idylla MSI Test and discontinue running the Idylla MSI Test on the listed Instrument(s). o No new Idylla MSI testing should be performed using the affected Idylla Instruments. o For results already obtained using an affected Idylla Instrument, the following apply: o MSS results are VALID and may be reported. o MSI-H (Microsatellite Instability High) results with six or seven positive identified biomarkers (as depicted in the result report) are VALID and may be reported. o MSI-H results other than those with six or seven positive identified biomarkers are INVALID and should NOT be reported or used to manage patient health. Patient samples with these results should be re-tested on an Idylla Instrument that is NOT affected. 2. If you are not the Idylla user, please forward this notice to the Idylla end user or Idylla supervisor and provide Biocartis with the correct contact details for our record keeping. 3. Completion of the Acknowledgement of Receipt (Appendix 2) is required for compliance. Please complete and sign the attached Acknowledgement of Receipt form by December 17, 2024, and send it to [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0836-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology