FDA RECALL RECORD · Z-0849-2025

8MM, LARGE NEEDLE DRIVER, IS4000 REF 470006 Recall

8MM, LARGE NEEDLE DRIVER, IS4000 REF 470006

FDA recall Class II

Ongoing

FDA lists this Class II recall for 8MM, LARGE NEEDLE DRIVER, IS4000 REF 470006. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0849-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 19, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
10,603
Distribution
See key details
FDA cause
Process change control

What is being recalled?

8MM, LARGE NEEDLE DRIVER, IS4000 REF 470006

Affected model & UDI checker

Search the exact identifiers published for Z-0849-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
4700068MM, LARGE NEEDLE DRIVER, IS4000 REF 470006Check source noticeNot specified
Not specified8MM, LARGE NEEDLE DRIVER, IS4000 REF 470006Check source notice00886874112151
FDA code information

Model Number: 470006 Part Number: 470006-12 UDI-DI code: 00886874112151 Batch Numbers: K10230120 K10230118 K12230119 K11230118 K10230217 Part Number: 470006-13 UDI-DI code: 00886874112151 Batch Numbers: K10230623 K10240506 K10240815

View the original FDA recall notice

Why was this product recalled?

Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

What actions did Intuitive Surgical, Inc. take?

On 12/19/2024, the firm send via email an "Urgent: Medical Device Correction" Letter informing customer that Intuitive Surgical has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci X and Xi reusable instruments. Customer are instructed to: 1. Customers can continue using the products in accordance with the user manual. 2. As a reminder, when using da Vinci X and Xi reusable instruments, follow the Inspection Before Use and Warnings listed in the manual provided with your system to inspect for any broken cables. Refer to Figures C and D for examples. In addition, please refer to Appendix B for additional images for detection of grip cable failures. 3. If you observe any failed (frayed or broken) grip cables prior to use, during procedure, or during reprocessing, please stop use of instrument, remove from use and inform Intuitive via the standard complaint process. 4. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject device via the standard complaint process. For Question and assistance - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history