8MM,LARGE HEM-O-LOK CLIP APPLIER,IS4000 REF 470230 Recall

8MM,LARGE HEM-O-LOK CLIP APPLIER,IS4000¿ REF 470230

FDA recall Class II

Ongoing

FDA lists this Class II recall for 8MM,LARGE HEM-O-LOK CLIP APPLIER,IS4000 REF 470230. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0861-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 19, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
29,559
Distribution
See key details
FDA cause
Process change control

What is being recalled?

8MM,LARGE HEM-O-LOK CLIP APPLIER,IS4000 REF 470230

Affected model & UDI checker

Search the exact identifiers published for Z-0861-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
4702308MM,LARGE HEM-O-LOK CLIP APPLIER,IS4000 REF 470230Check source noticeNot specified
Not specified8MM,LARGE HEM-O-LOK CLIP APPLIER,IS4000 REF 470230Check source notice00886874112380
FDA code information

Model Number: 470230 Part Number: 470230-12 UDI-DI code: 00886874112380 Batch Numbers: K10230103 K10230122 K10230123 K10230212 K10230226 K10230302 K10230310 K10230315 K10230323 K10230405 K10230420 K10230427 K10230504 K10230511 K10230518 K10230525 K10230602 K10230608 K10230615 K10230622 K10230706 K10230713 K10230727 K10230803 K10230810 K10230817 K10230921 K10231005 K10231012 K10231019 K10231027 K10231109 K10231123 K10231130 K10231207 K10231214 K10231218 K10240119 K10240125 K10240208 K10240215 K10240222 K10240229 K10240307 K10240327 K10240404 K10240411 K10240418 K10240516 K10240523 K10240606 K10240620 K10240627 K10240702 K10240711 K10240718 K10240808 K10240822 K11230103 K11230108 K11230212 K11230302 K11230315 K11230511 K11230518 K11230615 K11230622 K11230727 K11230817 K11230921 K11231005 K11231123 K11231214 K11240119 K11240125 K11240229 K11240404 K11240523 K11240620 K12230103 K12230108 K12230315 K12231116 K12240125 K12240404 K13230108 Model Number: 470230 Part Number: 470230-14 UDI-DI code: 00886874112380 Batch Numbers: K10230914

View the original FDA recall notice

Why was this product recalled?

Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

What actions did Intuitive Surgical, Inc. take?

On 12/19/2024, the firm send via email an "Urgent: Medical Device Correction" Letter informing customer that Intuitive Surgical has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci X and Xi reusable instruments. Customer are instructed to: 1. Customers can continue using the products in accordance with the user manual. 2. As a reminder, when using da Vinci X and Xi reusable instruments, follow the Inspection Before Use and Warnings listed in the manual provided with your system to inspect for any broken cables. Refer to Figures C and D for examples. In addition, please refer to Appendix B for additional images for detection of grip cable failures. 3. If you observe any failed (frayed or broken) grip cables prior to use, during procedure, or during reprocessing, please stop use of instrument, remove from use and inform Intuitive via the standard complaint process. 4. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject device via the standard complaint process. For Question and assistance - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology