FDA RECALL RECORD · Z-0866-2026

IntelliVue Patient Monitor MX600. Product Number: 865242. Recall

IntelliVue Patient Monitor MX600. Product Number: 865242.

FDA recall Class II

Ongoing

FDA lists this Class II recall for IntelliVue Patient Monitor MX600. Product Number: 865242.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0866-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 31, 2025
FDA report date
Dec 10, 2025
Quantity in commerce
1,913,441 units
Distribution
See key details
FDA cause
Use error

What is being recalled?

IntelliVue Patient Monitor MX600. Product Number: 865242.

Affected model & UDI checker

Search the exact identifiers published for Z-0866-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedIntelliVue Patient Monitor MX600. Product Number: 865242.Check source notice00884838029095
FDA code information

Product Number: 865242; UDI-DI: 00884838029095; All Serial No.

View the original FDA recall notice

Why was this product recalled?

Potential issue where the IntelliVue monitors did not alarm.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America Llc take?

An URGENT Medical Device Correction notice, dated 10/31/25, was mailed to customers and distributors. Distributors were notified of this correction and instructed to forward the provided recall notification to all consignees. Additionally, distributors are asked to include their contact information for consignees in the provided response form and perform a good faith effort for each customer to obtain a completed response form. Consignees are instructed to notify all users of this recall notification and forward the notice to all who need to be aware within their organization or where affected devices were potentially transferred. Consignees are to re-evaluate and reconfigure device configuration per the provided instructions. Once recall actions have been completed, consignees are to complete the provided response form either via QR code, by email to [email protected], or by fax to 877-499-7223. Philips is planning to develop a software update which will enforce the additional confirmation step when the user enables infinite Alarms Off for IntelliVue MP5 models only. Consignees with any questions can contact Philips at 800-722-9377.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology