FDA RECALL RECORD · Z-0876-2025

8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 471172 Recall

8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 471172

FDA recall Class II

Ongoing

FDA lists this Class II recall for 8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 471172. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0876-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 19, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
68,458
Distribution
See key details
FDA cause
Process change control

What is being recalled?

8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 471172

Affected model & UDI checker

Search the exact identifiers published for Z-0876-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
4711728MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 471172Check source noticeNot specified
Not specified8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 471172Check source notice00886874119792
FDA code information

Model Number: 471172 Part Number: 471172-16 UDI-DI code: 00886874119792 Batch Numbers: K10230831 K11230831 K12230831 K16240508 K17240508 K18240508 Model Number: 471172 Part Number: 471172-17 UDI-DI code: 00886874119792 Batch Numbers: K10221030 K10221122 K10221128 K10221212 K10221218 K10230103 K10230122 K10230129 K10230201 K10230205 K10230212 K10230219 K10230226 K10230302 K10230427 K10230504 K10230510 K10230515 K10230518 K10230526 K10230601 K10230608 K10230615 K10230622 K10230629 K10230706 K10230713 K10230714 K10230720 K10230727 K10230803 K10230810 K10230817 K10230824 K10230907 K10230914 K10230921 K10230928 K10231005 K10231012 K10231027 K10231109 K10231123 K10231201 K10231213 K10231214 K10231218 K10240104 K10240111 K10240118 K10240120 K10240125 K10240201 K10240208 K10240215 K10240229 K10240314 K10240327 K10240404 K10240410 K10240418 K10240425 K10240502 K10240515 K10240516 K10240523 K10240530 K10240603 K10240606 K10240613 K10240618 K10240702 K10240703 K10240708 K11221030 K11221212 K11221218 K11230122 K11230205 K11230302 K11230310 K11230504 K11230515 K11230518 K11230526 K11230601 K11230608 K11230615 K11230622 K11230629 K11230706 K11230713 K11230727 K11230803 K11230810 K11230817 K11230824 K11230907 K11230914 K11230921 K11230928 K11231005 K11231019 K11231102 K11231109 K11231123 K11231201 K11231214 K11240104 K11240111 K11240118 K11240125 K11240201 K11240215 K11240229 K11240404 K11240418 K11240502 K11240509 K11240523 K11240530 K11240606 K11240613 K11240618 K11240702 K11240711 K12221218 K12230122 K12230205 K12230515 K12230518 K12230601 K12230608 K12230615 K12230622 K12230629 K12230727 K12230803 K12230810 K12231005 K12231019 K12231214 K12240111 K12240201 K12240404 K12240502 K12240509 K12240523 K12240530 K12240711 K13230122 K13230515 K13230518 K13230601 K13230622 K13230810 K13231005 K13231019 K13240404 K13240509 K13240711 K14230122 K14230515 K14230518 K14230525 K15230122 K15230515 K15230525

View the original FDA recall notice

Why was this product recalled?

Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

What actions did Intuitive Surgical, Inc. take?

On 12/19/2024, the firm send via email an "Urgent: Medical Device Correction" Letter informing customer that Intuitive Surgical has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci X and Xi reusable instruments. Customer are instructed to: 1. Customers can continue using the products in accordance with the user manual. 2. As a reminder, when using da Vinci X and Xi reusable instruments, follow the Inspection Before Use and Warnings listed in the manual provided with your system to inspect for any broken cables. Refer to Figures C and D for examples. In addition, please refer to Appendix B for additional images for detection of grip cable failures. 3. If you observe any failed (frayed or broken) grip cables prior to use, during procedure, or during reprocessing, please stop use of instrument, remove from use and inform Intuitive via the standard complaint process. 4. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject device via the standard complaint process. For Question and assistance - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology