Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0885-2026

Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients. Recall

Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0885-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Nov 14, 2025
FDA report date
Dec 24, 2025
Quantity in commerce
30 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.

Affected model & UDI checker

Search the exact identifiers published for Z-0885-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedIvenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.Check source notice00811505030122
FDA code information

Software Versions 5.10.1 and prior; UDI: 00811505030122.

View the original FDA recall notice

Why was this product recalled?

Software version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Fresenius Kabi USA, LLC take?

Consignees were emailed an "URGENT - Medical Device Field Correction" notice dated 11/14/25. The notice instructs consignees to review and become familiar with the provided notification's risk mitigation actions; the notice is to be forwarded if affected units were further distributed and users are to be notified. Consignees are to facilitate the installation of the new Ivenix Infusion Management System (version 5.2.2) and Large Volume Pump (5.10.2) software versions utilizing the provided instructions. Consignees are to note that the LVP will not be available for use during the update. Completed Customer Reply Forms may be sent via email to [email protected]. Customer questions can be directed to Customer Support via email at [email protected] or by phone at 1-855-354-6387.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.