Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0890-2026

Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices. Recall

Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0890-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 10, 2025
FDA report date
Dec 17, 2025
Quantity in commerce
18 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.

Affected model & UDI checker

Search the exact identifiers published for Z-0890-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedEndo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.Check source notice04026575175222
FDA code information

Item Number: 15-0028/08; UDI-DI: 04026575175222; Lot Number: 2535004;

View the original FDA recall notice

Why was this product recalled?

The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.

What actions did Waldemar Link GmbH & Co. KG (Mfg Site) take?

On November 10, 2025, URGENT: MEDICAL DEVICE RECALL letters were emailed to distributors. Actions to be taken: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or [email protected] for any questions.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology