FDA RECALL RECORD · Z-0895-2025

GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission Recall

GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0895-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 18, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
8 units
Distribution
See key details
FDA cause
Package design/selection

What is being recalled?

GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission

Affected model & UDI checker

Search the exact identifiers published for Z-0895-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedGE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, EmissionCheck source notice840682121446
FDA code information

1) H3906AB: UDI/DI 840682121446, System ID UX2305NU03, Serial Number CZTV29005; UDI/DI 840682121446, System ID M4166103, Serial Number CZTV29006; UDI/DI 840682121446, System ID 00243NUC22, Serial Number CZTW29008; UDI/DI 840682121446, System ID 00222NUC07, Serial Number CZTW29009; UDI/DI N/A, System ID 604875D670B, Serial Number CZTW29007; UDI/DI N/A, System ID NT6200, Serial Number AD7W38007; UDI/DI N/A, System ID NT6201, Serial Number AD7X38017; UDI/DI N/A, System ID NT6100, Serial Number AP7W37001.

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/18/2024. The notice explained the safety issue and requested the following actions be taken: There are two populations potentially affected by this notification: NOTE: Specific serial numbers of the following affected systems: Infinia Nuclear Medicine Systems: Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4; VG Nuclear Medicine systems: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH; Brivo NM 615; Discovery NM 630; Discovery NM/CT 670, Discovery NM/CT 670 ES, Discovery NM/CT 670 DR, Discovery NM/CT 670 PRO, Discovery NM/CT 670 CZT; Optima NM/CT 640. For Serial Numbers listed in Appendix A of this letter: GE HealthCare requests that you STOP USING the system until GE HealthCare has inspected and if necessary, corrected your system and provided you with a letter indicating the system can be returned to clinical use. GE HealthCare has determined that these systems could have previously been transported or relocated without proper detector support. For Serial Numbers listed in Appendix B of this letter: You can continue to use your system, as GE HealthCare has determined that these systems have not previously been transported or relocated without proper detector support. For Serial Numbers listed in either Appendix A or Appendix B of this letter: GE HealthCare requests that you DO NOT physically relocate or transport your system from its current location. If you need to move or transport your system, please email GE HealthCare at [email protected] to receive specific instructions to follow before moving the system. When emailing this address, please provide the following information: " System serial number " Current system location " The reason for the move or transport, and subsequent destination Do not proceed with a physical relocation or transport of your system until you have received these instructions and taken the actions descri

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology