GE Healthcare Varicam Millennium VG Discovery VH, Model/Catalog Numbers: 1) ASM000085; 2) ASM001092; 3) SYS000001; System, Tomography, Computed, Emission Recall

GE Healthcare Varicam Millennium VG Discovery VH, Model/Catalog Numbers: 1) ASM000085; 2) ASM001092; 3) SYS000001; System, Tomography, Computed, Emission

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE Healthcare Varicam Millennium VG Discovery VH, Model/Catalog Numbers: 1) ASM000085; 2) ASM001092; 3) SYS000001; System, Tomography, Computed, Emission. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0900-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 18, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
30 units
Distribution
See key details
FDA cause
Package design/selection

What is being recalled?

GE Healthcare Varicam Millennium VG Discovery VH, Model/Catalog Numbers: 1) ASM000085; 2) ASM001092; 3) SYS000001; System, Tomography, Computed, Emission

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

1) ASM000085: UDI/DI N/A, System ID NM10569, Serial Number 10569; UDI/DI N/A, System ID NM10570, Serial Number 10570; UDI/DI N/A, System ID YN0845, Serial Number 10242; UDI/DI N/A, System ID YN0498, Serial Number 1092; UDI/DI N/A, System ID YN0440, Serial Number 1003; UDI/DI N/A, System ID 979764VGH1, Serial Number 77521848; UDI/DI N/A, System ID NM10225, Serial Number 10225. 2) ASM001092: UDI/DI N/A, System ID 310231VGH1, Serial Number 10539; UDI/DI N/A, System ID 207795VH, Serial Number 10571; UDI/DI N/A, System ID 82426070015, Serial Number 10550; UDI/DI N/A, System ID 82426210010, Serial Number 10595; UDI/DI N/A, System ID 82426280001, Serial Number 10591; UDI/DI N/A, System ID 082426080009, Serial Number 10599; UDI/DI N/A, System ID H2103NUC06, Serial Number 10582; UDI/DI N/A, System ID H1404NUC06, Serial Number 10604; UDI/DI N/A, System ID H2094NUC07, Serial Number 10587; UDI/DI N/A, System ID YN0984, Serial Number 10543; UDI/DI N/A, System ID 850260054, Serial Number 10579; UDI/DI N/A, System ID RU1591NM01, Serial Number 10561; UDI/DI N/A, System ID RU1457NM01, Serial Number 10553; UDI/DI N/A, System ID 209526MILVG, Serial Number 10340; UDI/DI N/A, System ID 416469DVH, Serial Number 10384; UDI/DI N/A, System ID CY1088MN04, Serial Number 10360; UDI/DI N/A, System ID Not Available, Serial Number 81251648; UDI/DI N/A, System ID B5315325, Serial Number 10185. 3) ASM001099: UDI/DI N/A, System ID PER10925, Serial Number 10346. 4) SYS000001: UDI/DI N/A, System ID NM10374, Serial Number 10374; UDI/DI N/A, System ID NM10147, Serial Number 10147; UDI/DI N/A, System ID 334794MILVG, Serial Number 1063; UDI/DI N/A, System ID NM10291, Serial number 10291; UDI/DI N/A, System ID 56948, Serial number 1021. 5) SYS000002: UDI/DI N/A, System ID 949760VG1, Serial Number 10445; UDI/DI N/A, System ID NM10174, Serial Number 10174; UDI/DI N/A, System ID 223410MILVG, Serial Number 1069. 6) SYS000003: UDI/DI N/A, System ID 914734VG, Serial number 10435.

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/18/2024. The notice explained the safety issue and requested the following actions be taken: There are two populations potentially affected by this notification: NOTE: Specific serial numbers of the following affected systems: Infinia Nuclear Medicine Systems: Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4; VG Nuclear Medicine systems: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH; Brivo NM 615; Discovery NM 630; Discovery NM/CT 670, Discovery NM/CT 670 ES, Discovery NM/CT 670 DR, Discovery NM/CT 670 PRO, Discovery NM/CT 670 CZT; Optima NM/CT 640. For Serial Numbers listed in Appendix A of this letter: GE HealthCare requests that you STOP USING the system until GE HealthCare has inspected and if necessary, corrected your system and provided you with a letter indicating the system can be returned to clinical use. GE HealthCare has determined that these systems could have previously been transported or relocated without proper detector support. For Serial Numbers listed in Appendix B of this letter: You can continue to use your system, as GE HealthCare has determined that these systems have not previously been transported or relocated without proper detector support. For Serial Numbers listed in either Appendix A or Appendix B of this letter: GE HealthCare requests that you DO NOT physically relocate or transport your system from its current location. If you need to move or transport your system, please email GE HealthCare at [email protected] to receive specific instructions to follow before moving the system. When emailing this address, please provide the following information: " System serial number " Current system location " The reason for the move or transport, and subsequent destination Do not proceed with a physical relocation or transport of your system until you have received these instructions and taken the actions descri

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology