Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Recall
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support. Digoxin is a widely prescribed cardiac glycoside indicated in the treatment of congestive heart failure and supraventricular arrhythmias. Digoxin measurements are used to monitor patient compliance and therapy, and to diagnose potential overdose. VITROS Chemistry Products DGXN Slides quantitatively measure digoxin (DGXN) concentration in serum using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. One sales unit contains 5 cartridges. Each cartridge holds 18 slides (equivalent to one test), for 90 slides per sales unit. Component: No, not a component
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0902-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 5, 2024
- FDA report date
- Jan 22, 2025
- Quantity in commerce
- 1,541 units (1,125 US, 416 OUS)
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support. Digoxin is a widely prescribed cardiac glycoside indicated in the treatment of congestive heart failure and supraventricular arrhythmias. Digoxin measurements are used to monitor patient compliance and therapy, and to diagnose potential overdose. VITROS Chemistry Products DGXN Slides quantitatively measure digoxin (DGXN) concentration in serum using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. One sales unit contains 5 cartridges. Each cartridge holds 18 slides (equivalent to one test), for 90 slides per sales unit. Component: No, not a component
Affected model & UDI checker
Search the exact identifiers published for Z-0902-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 8343386 | Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides | Check source notice | Not specified |
| Not specified | Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides | Check source notice | 10758750004782 |
FDA code information
Model No 8343386; UDI/GTIN 10758750004782; Coating 0272, expiration date; 01-April-2025 The scope of this recall is limited to consignees who received VITROS Chemistry Products DGXN slides in the last 18 months. This will include in-date lots manufactured from coatings (production formulation) 0271 and/or 0272. Of the in date product, the latest expiration date is listed for the coating.
Why was this product recalled?
Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results.
What actions did QUIDEL ORTHO take?
On December 5, 2024 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to consignees. REQUIRED ACTIONS - For customers who have VITROS 5,1 FS/4600/5600/XT 3400/XT 7600 Systems: Load ADD DRV 6308 upon receipt. - For customers who have VITROS 250/350 Systems: Please be aware of the lowering of the Interferent Concentration limit from 300 mg/dL to 100 mg/dL at a digoxin concentration of 2.0 ng/mL (2.6 nmol/L). - Complete and return the enclosed Confirmation of Receipt form no later than December 13, 2024. - Save this notification with your User Documentation or post this notification by each VITROS 250/350/5,1 FS/4600/5600/XT 3400/XT 7600 System in your laboratory until the issue has been resolved. - Please forward this notification if the affected product was distributed outside of your facility. - If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. Resolution Our investigation to revise the limit at which hemoglobin interference may reach an unacceptable level is ongoing. In the interim we will be issuing a Technical Bulletin to implement the temporary changes to the Sample Indices Threshold default hemolysis limit on the ADD and the hemoglobin Interferent Concentration limit for digoxin listed in the IFU. We will communicate again after our investigation has completed. QuidelOrtho is targeting to have an update by Q1 2025. Contact Information We apologize for the inconvenience this may cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-421-3311. To report adverse events, contact your local Global Services Organization. Adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adverse Event Reporting program: www.fda.gov/medwatch/report.htm, 1-800-332-1088. On December 5, 2024 URGENT PRODUCT CORRECTION NOTIFICATION lette
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0902-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology