Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Recall
Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type).. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0905-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Dec 18, 2024
- FDA report date
- Jan 22, 2025
- Quantity in commerce
- 30,140 units
- Distribution
- See key details
- FDA cause
- Labeling design
What is being recalled?
Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.
Affected model & UDI checker
Search the exact identifiers published for Z-0905-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MAJ-891 | Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). | Check source notice | Not specified |
| Not specified | Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). | Check source notice | 04953170063114 |
FDA code information
Model Number: MAJ-891. UDI-DI: 04953170063114. All lot numbers.
Why was this product recalled?
Potential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Olympus Corporation of the Americas take?
Olympus notified consignees on about 12/18/2024 via letter. Consignees were instructed to utilize alternative devices to the MAJ-891 according to the provided compatibility table provided, ensure all personnel are completely knowledgeable and thoroughly trained on the content of this letter and the MAJ-891 reprocessing instructions, acknowledge receipt of the letter through the Olympus portal, https://olympusamerica.com/recall, and notify customers if the units have been further distributed. Additionally, Olympus will contact customers to schedule a mutually beneficial time for an Olympus Field Representative to visit each facility and review the contents of this letter and the MAJ-891 reprocessing instructions.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0905-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology