BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352 Recall
BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352
FDA recall Class II
Ongoing
FDA lists this Class II recall for BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0906-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 5, 2023
- FDA report date
- Jan 22, 2025
- Quantity in commerce
- 622
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352
Affected model & UDI checker
Search the exact identifiers published for Z-0906-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 323 | BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352 | Check source notice | Not specified |
| 352 | BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352 | Check source notice | Not specified |
FDA code information
Software v1.7.3. REF/UDI-DI: 323/10885403512667, 352/10885403512674
Why was this product recalled?
Due to an automated dispensing cabinet software issue, when a medication single order contains more than one component and at least one is unavailable for a multi-item or combination order, the order may display as grayed out; alternate location tooltip will display "Server connection failed," not listing other area devices where unavailable item(s) may be found, which may lead to delayed therapy.
What actions did CareFusion 303, Inc. take?
On 4/5/2023, correction notices were sent to customers who were asked to do the following: A permission-based feature called "Global Find" can be used to obtain alternate storage locations of medication(s) not loaded (by pharmacy) or depleted at that device and is not impacted by this issue. 1) Review and confirm that backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them to avoid delays in access to medications. 2) Ensure this letter is forwarded to applicable facilities if devices have been transferred and that all appropriate personnel have been made aware of this issue. 3) Complete and return the enclosed Customer Response Form via email to [email protected] 4) Adverse health consequences experienced with the use of this product should be reported to BD at 1-800-727-6102 or via bd.com/self-service. Recall/Letter Questions: BD Customer Support, Phone: 1-866-583-8783, Phone Hours: 5:00am PST to 4:00pm PST (8:00am ET to 7:00pm ET), Monday - Friday A software update is available. BD will contact you to schedule a time to install the software update.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0906-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology