Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0907-2025

Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD. Recall

Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0907-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 13, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
380
Distribution
See key details
FDA cause
Process control

What is being recalled?

Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.

Affected model & UDI checker

Search the exact identifiers published for Z-0907-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PRD-04484Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.Check source noticeNot specified
Not specifiedPanther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.Check source notice15420045510890
FDA code information

UDI-DI: 15420045510890, Kit Lot 715927, containing Cartridge Lot 624907, Exp. Date 2025-08-15.

View the original FDA recall notice

Why was this product recalled?

Potential weak pouch seal on Panther Fusion GBS assay cartridge, causing potential for invalid or incorrect results leading to a possible delay of treatment.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Hologic, Inc. take?

An Urgent Medical Device Recall Notification letter issued 12/13/2024 via Fed Ex. The letter instructs customers to discontinue use of the Panther Fusion GBS Assay kit lot 715927 containing cartridge lot 624907 and segregate and count the number of affected Panther Fusion GBS Assay kits from lot 715927 containing cartridge lot 624907. Customers are instructed to destroy the product and fill out the online Customer Response Form (CRF) indicating that the on-hand inventory of recalled Panther Fusion GBS Assay Kit has been destroyed. If you have any questions or concerns regarding this recall notification, please contact Hologic Technical Support at +1 888 484 4747 or +1 858 410 8511 during the hours of 5 am PT and 5 pm PT, or by e-mail at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.