FDA lists this Class II recall for DEXLOCK Achilles Repair Implant Kits, MAKT4520. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0911-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 11, 2025
FDA report date
Dec 17, 2025
Quantity in commerce
163 US
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

DEXLOCK Achilles Repair Implant Kits, MAKT4520

Affected model & UDI checker

Search the exact identifiers published for Z-0911-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedDEXLOCK Achilles Repair Implant Kits, MAKT4520Check source notice10193489131703
FDA code information

Lot 19019 UDI-DI 10193489131703

View the original FDA recall notice

Why was this product recalled?

There have been multiple complaints of drill bit fusing to the bushing/sleeve within the drill guide during use. Retrieving a replacement drill bit and drill guide may prolong the surgical procedure.

What actions did Medline Industries, LP take?

Firm began notifying consignees on 11/11/2025 via Urgent Medical Device Recall letters. Customers have been instructed to over-label kits with a sticker instructing them to not use the drill guide within the Achilles Kit and to use the reusable 3.6mm drill guide from the Soft Tissue Instrument Tray. Customers who do not have a Soft Tissue Instrument Tray have been instructed to notify Medline so that a tray can be provided. Customers were instructed to quarantine all affected product in stock and return a completed response form to Medline. Upon receipt of the response form, the firm will then provide over-labels to place on affected inventory. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology