Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0915-2026

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S Recall

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

FDA recall Class I

Ongoing

FDA lists this Class I recall for GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0915-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Nov 14, 2025
FDA report date
Dec 24, 2025
Quantity in commerce
203 units
Distribution
See key details
FDA cause
Component change control

What is being recalled?

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

FRU 2076139-001-S, distributed between June 2, 2025 and October 21, 2025

View the original FDA recall notice

Why was this product recalled?

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Medical Systems China Co., Ltd. take?

GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 11/14/2025 via letter using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken by Customer/User: "Pending correction by GE HealthCare, you can continue to use the Anesthesia machine with affected power management boards by following the instructions below: 1. Always ensure the device has a secure connection to an AC mains power source. 2. If there is a loss of AC mains power to the system leading to an unexpected system shutdown: - Promptly initiate ventilation using a self-inflating bag connected to an oxygen source. - Assess oxygenation via pulse oximetry. - Because volatile anesthetic agent delivery may transiently be disrupted, supplement with or transition to intravenous anesthetics as needed. - Following system reboot, the system will enter pre-use check. Press "Start Anesthesia or Start Case and then select the "Bypass" button to bypass the checkout. Proceed to selecting the ventilation parameters and volatile agent concentration appropriate for the patient. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to [email protected] or submit the acknowledgment using survey." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology