Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0921-2026

SCISSORS, OR, SHARP/BLUNT, 5.5", STERILE, Item Number DYND04000 Recall

SCISSORS, OR, SHARP/BLUNT, 5.5", STERILE, Item Number DYND04000

FDA recall Class II

Ongoing

FDA lists this Class II recall for SCISSORS, OR, SHARP/BLUNT, 5.5", STERILE, Item Number DYND04000. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0921-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 13, 2025
FDA report date
Dec 24, 2025
Quantity in commerce
56 units
Distribution
See key details
FDA cause
Packaging process control

What is being recalled?

SCISSORS, OR, SHARP/BLUNT, 5.5", STERILE, Item Number DYND04000

Affected model & UDI checker

Search the exact identifiers published for Z-0921-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedSCISSORS, OR, SHARP/BLUNT, 5.5", STERILE, Item Number DYND04000Check source notice10193489191349
Not specifiedSCISSORS, OR, SHARP/BLUNT, 5.5", STERILE, Item Number DYND04000Check source notice20193489191346
FDA code information

DYND04000, UDI-DI: 10193489191349(each), 20193489191346(case), Lot Number: 2023102590

View the original FDA recall notice

Why was this product recalled?

The items may have weak packaging seals that may result in a breach of sterility if the seal fails and may lead to an increased risk of infection if product is utilized.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medline Industries, LP take?

An URGENT MEDICAL DEVICE RECALL notification letter dated 11/13/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed on the enclosed response form. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please complete and return the enclosed response form listing the quantity of affected product you have in inventory. Even if you do not have any affected product in inventory, please complete and return the response form. 3. Your account will receive credit once the response form is submitted, if applicable. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology