Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0922-2025

ARTIS One Angiographic X-Ray System Recall

ARTIS One Angiographic X-Ray System

FDA recall Class II

Ongoing

FDA lists this Class II recall for ARTIS One Angiographic X-Ray System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0922-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 2, 2025
FDA report date
Feb 5, 2025
Quantity in commerce
1140
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

ARTIS One Angiographic X-Ray System

Affected model & UDI checker

Search the exact identifiers published for Z-0922-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
10848600ARTIS One Angiographic X-Ray SystemCheck source noticeNot specified
FDA code information

Model Number 10848600

View the original FDA recall notice

Why was this product recalled?

A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Medical Solutions USA, Inc. take?

Manufacturer action instructions are not included in the retrieved record. Consult the original FDA recall notice for the instructions that apply to the affected product.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA classification recorded

  3. FDA notice posted

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.