FDA RECALL RECORD · Z-0940-2026

MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100) Recall

MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)

FDA recall Class II

Ongoing

FDA lists this Class II recall for MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0940-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 6, 2025
FDA report date
Dec 24, 2025
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
Component change control

What is being recalled?

MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)

Affected model & UDI checker

Search the exact identifiers published for Z-0940-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)Check source notice04260223141355
FDA code information

Part Number: 8-69-41100; UDI-DI: 04260223141355; Serial Numbers: 1008852, 1008888, 1008916, 1008925, 1008935, 1008945, 1008951, 1008957, 1008960, 1008962, 1008966, 1008967, 1008987, 1008990, 1009054, 1009065, 1009066, 1009131, 1009137, 1009138, 1009139, 1009145, 1009148, 1009155, 1009159, 1009161, 1009169, 1009184, 1009188, 1009191, 1009305, 1009315, 1009316, 1009319, 1009321, 1009322, 1009325, 1009328, 1009343, 1009352, 1009356, 1009357, 1009359, 1009366, 1009372, 1009375, 1009376, 1009386, 1009388, 1009390, 1009396, 1009399, 1009400, 1009404, 1009411, 1009415, 1009419, 1009426, 1009427, 1009432, 1009439, 1009442, 1009454, 1009456, 1009460, 1009463, 1009464, 1009468, 1009475, 1009481, 1009488, 1009494, 1009495, 1009497, 1009848, 1009855.

View the original FDA recall notice

Why was this product recalled?

Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did PATH take?

On November 6, 2025 URGENT: MEDICAL DEVICE Field Corrective Action letters were sent to their two distributors. Actions to be taken: - Users will be asked to confirm the DPOAE protocol in use. - Verify the DPOAE Protocol used is not affected by the issue (DP1, DP2, DP3, DP4, DP17, DP18, DP19, DP20, DP21). - If any other protocol is used, please replace it by using AccuLink Software. Follow instructions in IFU to do so. - Remove all affected protocols (DP5-DP16) using AccuLink Software. Follow instructions in IFU to do so. - Complete and return the attached form to confirm these steps were followed. - The expected lifetime of the probes is two years from date of manufacture. Please put the products out of use, once this is reached.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.