Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0958-2026

InPen App, Model/CFN Number: MMT-8061 (Android Users) Recall

InPen App, Model/CFN Number: MMT-8061 (Android Users)

FDA recall Class II

Ongoing

FDA lists this Class II recall for InPen App, Model/CFN Number: MMT-8061 (Android Users). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0958-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 13, 2025
FDA report date
Dec 24, 2025
Quantity in commerce
8251
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

InPen App, Model/CFN Number: MMT-8061 (Android Users)

Affected model & UDI checker

Search the exact identifiers published for Z-0958-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedInPen App, Model/CFN Number: MMT-8061 (Android Users)Check source notice00763000974596
FDA code information

Software Versions: 7.5.0, 7.5.1, and 8.0.0. UDI-DI: 00763000974596

View the original FDA recall notice

Why was this product recalled?

When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.

What actions did Medtronic MiniMed, Inc. take?

On 11/13/2025, correction notices were emailed to customers who were asked to update to version 8.0.1 or higher as soon as possible to resolve this issue. 1.Verify your app version: -On the home screen, navigate to the Settings menu (lower right corner). - Select Help and Support. - Note the App Version displayed. 2. If you are running versions 8.0.0, update to the latest version 8.0.1 or higher via the Google Play Store: - Open the Google Play Store. - Tap your profile icon at the top of the screen, then select Manage apps & device. - Tap Updates available or search for the InPen app. - Tap Update next to the InPen app icon. Electronically acknowledge that you have read and understand this notification and have followed the actions listed For questions, contact the firm's 24-Hour Technical Support line at 1-800-646-4633, option #1. The Customer Confirmation Form can also be emailed to the firm at [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology