FDA RECALL RECORD · Z-0960-2026

Brand Name: Xpert BCR-ABL Ultra Recall

Brand Name: Xpert¿ BCR-ABL Ultra Model/Catalog Number: GXBCRABL-10

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Xpert BCR-ABL Ultra. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0960-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 30, 2025
FDA report date
Dec 24, 2025
Quantity in commerce
4032 units
Distribution
See key details
FDA cause
Process design

What is being recalled?

Brand Name: Xpert BCR-ABL Ultra Model/Catalog Number: GXBCRABL-10

Affected model & UDI checker

Search the exact identifiers published for Z-0960-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
GXBCRABL-10Brand Name: Xpert BCR-ABL UltraCheck source noticeNot specified
FDA code information

UDI: 007332940001834/Lot Numbers: 1001452775 1001457465 1001457622 1001459091 1001462586 1001462625 1001463944 1001465391 1001469480 1001473229 1001474930 1001477474 1001480636 1001483324 1001486594 1001492908 1001493823 1001495811 1001498202 1001501650

View the original FDA recall notice

Why was this product recalled?

Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results.

What actions did Cepheid take?

On November 14, 2025, Cepheid issued a "Urgent Medical Device Recall" notification to affected consignees. Cepheid asked consignees to take the following actions: 1. Cepheid recommends that laboratories and clinical facilities utilizing the Xpert BCR-ABL Ultra assay for quantification of BCR::ABL1 and ABL1 mRNA transcript levels in peripheral blood specimens from t(9;22)-positive chronic myeloid leukemia (CML) patients undergoing tyrosine kinase inhibitor (TKI) therapy, assess white blood cell count (WBCC) prior to testing in cases where elevated white blood cell count is suspected. 2. If WBCCC exceeds 30,000 cells/uL, the specimen should be processed in the accordance with the Retest Procedure for ERROR (Code 2008) or INVALID (Type 2) as detailed in the IFU for the following scenarios: - Cases where INVALID test results are observed and are associated with an elevated WBCC. - Cases of valid results in which an elevated WBCC has been observed. 3. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. 4. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them with a copy of this letter. 5. Please complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology