FDA RECALL RECORD · Z-0964-2026

da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40 Recall

da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

FDA recall Class II

Ongoing

FDA lists this Class II recall for da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0964-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 21, 2024
FDA report date
Dec 31, 2025
Quantity in commerce
144
Distribution
See key details
FDA cause
Device Design

What is being recalled?

da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

Affected model & UDI checker

Search the exact identifiers published for Z-0964-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedda Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40Check source notice00886874119747
FDA code information

UDI-DI: 00886874119747. Serial Numbers: 10476314, 10485693, 10496049, 10497992, 10503432, 10503433, 10507302, 10509551, 10511044, 10511047, 10516467, 10518960, 10518961, 10522438, 10522439, 10522441, 10523771, 10523773, 10523775, 10523776, 10523777, 10523778, 10523779, 10523780, 10523781, 10529917, 10529918, 10529919, 10529920, 10529922, 10534426, 10534428, 10534429, 10534430, 10534752, 10534753, 10534754, 10534755, 10534757, 10543027, 10543028, 10543048, 10543051, 10544670, 10544676, 10544677, 10544678, 10547713, 10547717, 10547724, 10547728, 10551553, 10551554, 10552616, 10552618, 10552620, 10552621, 10556681, 10556682, 10558195, 10558197, 10558198, 10558202, 10560603, 10560604, 10560607, 10562584, 10562585, 10562586, 10562588, 10565609, 10565610, 10565614, 10567141, 10567143, 10567145, 10569365, 10569367, 10569368, 10569369, 10570704, 10570705, 10570706, 10570707, 10573677, 10573678, 10573679, 10573680, 10573681, 10573682, 10573683, 10573684, 10573685, 10573686, 10576019, 10576021, 10576022, 10576023, 10576024, 10576025, 10576026, 10576027, 10576028, 10579442, 10579446, 10579453, 10579455, 10579456, 10579457, 10579461, 10579466, 10582427, 10582428, 10582429, 10591177, 10591179, 10591180, 10591184, 10591197, 10594614, 10594615, 10594616, 10594617, 10594618, 10594619, 10597107, 10597108, 10597109, 10597110, 10599042, 10599044, 10599045, 10599048, 10600750, 10600753, 10603334, 10603335, 10603343, 10603344, 10603345, 10603367, 10605468, 10606129, 10606133

View the original FDA recall notice

Why was this product recalled?

Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Intuitive Surgical, Inc. take?

On 8/21/2024, correction notices were sent to customers who were informed of the following: 1) Share this notification with all personnel at your site who perform and support da Vinci procedures. 2) Inform affected personnel when the correction has been completed. 3) Complete and return the acknowledgement form via email to [email protected] 4) Retain a copy of this letter and the acknowledgement form for your files. 5) Firm is actively working on making updated motor module assemblies available. Once available, an Intuitive representative will schedule a visit to replace the affected motor module sub-assembly. If you need further information or support concerning this notification, please contact the firm's Customer Service: North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology