Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0966-2026

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02 Recall

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

FDA recall Class II

Ongoing

FDA lists this Class II recall for LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0966-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 24, 2025
FDA report date
Dec 31, 2025
Quantity in commerce
12 customers
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

Affected model & UDI checker

Search the exact identifiers published for Z-0966-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedLifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02Check source notice00887709127227
FDA code information

LifeShield Infusion Safety Software Suite v2.2 UDI-DI 00887709127227

View the original FDA recall notice

Why was this product recalled?

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Icu Medical, Inc. take?

On November 24, 2025, the firm notified customers via Urgent Medical Device Correction letters. Consignees were asked to take the following actions: 1. Ensure all users or potential users of LifeShield DLM 2.2.1 are immediately made aware of this notification and proposed mitigations. 2. Until a software update is released, limit any newly created or edited Concentration Limits to one decimal place of precision for limits less than 100. ICU Medical is developing a software update to correct this issue and will notify customers of the timing of the software update.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology