Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment. Recall
Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0987-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 6, 2024
- FDA report date
- Jan 29, 2025
- Quantity in commerce
- 25 systems
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment.
Affected model & UDI checker
Search the exact identifiers published for Z-0987-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MN6000 | Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment. | Check source notice | Not specified |
| CM3100 | Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment. | Check source notice | Not specified |
| Not specified | Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment. | Check source notice | 05414734510035 |
FDA code information
Merlin.net Patient Care Network Heart Failure Web Application Model Number: MN6000 UDI-DI code: 05414734510035 CardioMEMS Hospital system Model: CM3100 Serial Numbers: M220200018 M220200028 M220300070 M220300173 M220300186 M220700135 M220900036 M220900045 M220900050 M220900052 M220900089 M220900109 M221100017 M221100020 M221100043 M221100060 M230500037 M230500038 M230900026 M230900037 M231100022 M240100031 M240100086 M240500086 M240500088
Why was this product recalled?
Due to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.
What actions did St Jude Medical take?
On 11/06/2024, the firm sent an "URGENT: MEDICAL DEVICES CORRECTION", to customers to inform them that on September 30, 2024, the Merlin.net PCN HF Portal (MN6000) was migrated to the cloud database and following the migration, some customers reported that their Hospital Systems were unable to transmit to the Merlin.net PCN. User Action Requested: Abbott requests that clinicians take the following actions: The Impacted CM3100 Hospital System devices at your facility is/are listed in the enclosed Appendix. o For a Hospital System marked SESSIONS SENT no further correction is needed. o Hospital Systems marked UNSENT SESSIONS should be turned on to send unsent sessions: If your Hospital System has connectivity, it will immediately attempt to send after boot-up. A spinning icon will appear to indicate that transmission is in progress. If you see a button on the home screen showing X Unsent Sessions select the button and follow the prompts to send. If you are unable to send sessions, contact Abbott Remote Care Technical Support (RCTS) for assistance. Distribute this notice to appropriate personnel within your institution. Complete the included acknowledgment form and return it to Abbott. NOTE: Hospital Systems that were not used between September 30, 2024, and October 15, 2024, are not impacted. any questions about this notice or potential patient impact, please contact your Abbott representative or Remote Care Technical Support at 1-844-MYCMEMS (692-6367) Monday Friday, 5AM PST 5PM PST.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0987-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology