The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement. Recall
The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement.
FDA recall Class II
Ongoing
FDA lists this Class II recall for The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0998-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 27, 2025
- FDA report date
- Jan 14, 2026
- Quantity in commerce
- 56
- Distribution
- See key details
- FDA cause
- Radiation Control for Health and Safety Act
What is being recalled?
The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Part Number V000629
Why was this product recalled?
Vision RT is informing customers of an omission identified in the Instructions for Use for AlignRT InBore systems. Information identifying the lasers and their characteristics was omitted from the Instructions for Use for the subject system as required by 21 CFR 1040,10(h), via conformance with recognized consensus standard IEC 60825-1 Ed. 3 (FDA Laser Notice 56).
What actions did Vision RT Ltd take?
Vision RT wrote to inform customers of an omission that they have identified in the Instructions for Use for AlignRT InBore systems. This notification contained information for users of AlignRT InBore systems on the content omitted and how to access the corrected Instructions for Use.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA classification recorded
FDA notice posted
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0998-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology