LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471 Recall
LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471
FDA recall Class II
Ongoing
FDA lists this Class II recall for LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0999-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 5, 2025
- FDA report date
- Jan 14, 2026
- Quantity in commerce
- U.S.: 1, OUS: 95
- Distribution
- See key details
- FDA cause
- Radiation Control for Health and Safety Act
What is being recalled?
LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471
Affected model & UDI checker
Search the exact identifiers published for Z-0999-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 11574100 | LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471 | Check source notice | Not specified |
| 10762471 | LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471 | Check source notice | Not specified |
FDA code information
(01)04056869269931(21)10330
Why was this product recalled?
A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the collimation may be incorrect.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Siemens Medical Solutions USA, Inc. take?
The Customer Advisory Notice (CAN) was sent to the affected customer on December 5, 2025. The CAN notifies the customer of the problem and provides steps the user should take to avoid the problem until the resolution software update is available. When switching within five seconds from a table related OGP to a bucky wall stand related OGP with top-alignment and then back to a table related OGP, it may happen, that the collimation is not adapted to the limitations of the detector. The active radiation field may thus extend beyond the limits of the detector, resulting in radiation outside of the area of the detector.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0999-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026