Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Recall
Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/Catalog Number: 6802783 (New), 6802914 (Certified/Refurbished), 6904000 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: no, not a component
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1007-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 19, 2024
- FDA report date
- Jan 29, 2025
- Quantity in commerce
- 981 units (99 US, 882 OUS)
- Distribution
- See key details
- FDA cause
- Software Design Change
What is being recalled?
Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/Catalog Number: 6802783 (New), 6802914 (Certified/Refurbished), 6904000 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: no, not a component
Affected model & UDI checker
Search the exact identifiers published for Z-1007-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 6802914 | Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System | Check source notice | 10758750002979 |
| 6904000 | Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System | Check source notice | 10758750035779 |
| 6802783 | Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System | Check source notice | Not specified |
FDA code information
Lot Code: VITROS 3600 Immunodiagnostic System Product Code: 6802783 (New) UDI: 10758750002979 VITROS 3600 Immunodiagnostic System (Certified/Refurbished) Product Code: 6802914 UDI: 10758750002979 VITROS 3600 Immunodiagnostic System (China) Product Code: 6904000 UDI: 10758750035779 Range of installed serial numbers:36000004-36001776 Manufacture date range: June 2008 to present Distribution date range; June 2008 to present Total systems currently installed in the US: 99 Total systems currently installed outside the US: 882 Note1: The catalogue number 6802783 is used to denote the VITROS 3600 Immunodiagnostic System. Catalogue number 6802914 is used to denote the VITROS 3600 Immunodiagnostic System that have been refurbished, these systems can also be referred to as certified. A VITROS 3600 Immunodiagnostic System Certified/Refurbished maintains its original serial number through the refurbishment/certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today. Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned.
Why was this product recalled?
The VITROS 3600 Immunodiagnostic System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did ORTHO-CLINICAL DIAGNOSTICS, INC. take?
On December 19, 2024 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. REQUIRED ACTION " Do not use expired Signal Reagent packs. See the Questions & Answers section for more details. " Complete the enclosed Confirmation of Receipt form no later than December 31, 2024. " Save this notification with your User Documentation or post this notification by each VITROS 3600/5600/XT 7600 System in your laboratory until the issue has been resolved. Contact Information We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact Global Services Organization at 1-800-421-3311. To report adverse events, contact your local Global Services Organization. Adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adverse Event Reporting program: www.fda.gov/medwatch/report.htm, 1-800-332-1088.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1007-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology