Brand Name: N/A Product Name: Repeater Bone Tamp Recall
Brand Name: N/A Product Name: Repeater Bone Tamp Model/Catalog Number: ASY-00132 Software Version: N/A Product Description: The Repeater Bone Tamp is a cylindrical shaft used to push bone graft material through a corresponding bone funnel into the FlareHawk interbody fusion device in spinal fusion procedures. Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: N/A Product Name: Repeater Bone Tamp. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1008-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 20, 2024
- FDA report date
- Jan 29, 2025
- Quantity in commerce
- 9
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: N/A Product Name: Repeater Bone Tamp Model/Catalog Number: ASY-00132 Software Version: N/A Product Description: The Repeater Bone Tamp is a cylindrical shaft used to push bone graft material through a corresponding bone funnel into the FlareHawk interbody fusion device in spinal fusion procedures. Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-1008-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| ASY-00132 | Brand Name: N/A Product Name: Repeater Bone Tamp | Check source notice | Not specified |
| Not specified | Brand Name: N/A Product Name: Repeater Bone Tamp | Check source notice | 00810004721906 |
FDA code information
Lot Code: Part Number: ASY-00132 Lot Number: 03200045 UDI Number: (01)00810004721906(10)03200045
Why was this product recalled?
Due to incompatibility of bone tamp (Rev A) and bone funnels (Rev C).
What actions did Integrity Implants Inc. take?
On 12/20/2024, the firm emailed a "NOTICE OF RECALL" letter to customer informing them that the firm has received a report relating to a specific lot of Repeater Bone Tamps not fitting with newer lots of the Repeater Bone Funnel. Customers are instructed to: 1. Segregate Product. immediately remove all Repeater Bone Tamp (Part # ASY-00132, Lot Number 03200045), from their inventory (regardless of its location) and segregate this product in a secure location for return to Accelus. 2. Complete QF-00070 Product Tracking & Verification Form. Complete and return the enclosed QF-00070 Product Tracking & Verification Form (even if you do not have any product to return), following the directions on this page and QF-00070 Product Tracking & Verification Form. Accelus' customer service representative can assist in completing the form. 3. Indicate Returned Products on QF-00070 Product Tracking & Verification Form. QF-00070 Product Tracking & Verification Form lists the Repeater Bone Tamp and product descriptions. Mark the number of each type of Repeater Bone Tamp that will be returned on this form. 4. Call (561)529-3861 or 1-800-201-9300 from 8:30am to 5pm (EDT) should there be any questions about returning the product. Customer's Accelus account will be credited in an amount equal to the actual purchase price for the returned product, if applicable. Inform Customer Service Representative whether customers would like to order a replacement product. 5. Package the Return Material. Package Repeater Bone Tamp(s) that are being returned in an appropriate shipping box. Place a signed original copy of QF-00070 Product Tracking & Verification Form in the box, appropriately seal the box and send it on or before 12-26-2024, to: Accelus 354 Hiatt Drive, Suite 100 Palm Beach Gardens, FL 33418 Office: (561)529-3861 or 1-800-201-9300 Fax: (561) 529-3869
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1008-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.