FDA RECALL RECORD · Z-1008-2026

BVI Leos Laser and Endoscopy System. Recall

BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

FDA recall Class II

Ongoing

FDA lists this Class II recall for BVI Leos Laser and Endoscopy System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1008-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 5, 2025
FDA report date
Jan 7, 2026
Quantity in commerce
5 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

Affected model & UDI checker

Search the exact identifiers published for Z-1008-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
OME6000U.BVI Leos Laser and Endoscopy System.Check source notice00886158037576
OME6000UBVI Leos Laser and Endoscopy System.Check source noticeNot specified
FDA code information

Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.

View the original FDA recall notice

Why was this product recalled?

Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

What actions did Beaver-Visitec International, Inc. take?

Beaver-Visitec International (BVI) notified consignees via emailed letter on 12/05/2025. Consignees were informed that the affected units can still be used. A BVI service technician will contact each consignee to schedule a visit. During this visit, the missing ferrites will be installed to restore the equipment to its fully compliant and validated configuration. BVI aims to complete the correction by December 31st, 2025.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology