Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1010-2026

Kii Low Profile, Bladed Dual Pack , Qty. 1 5x55mm REF: CTB23 Recall

Kii Low Profile, Bladed Dual Pack , Qty. 1 5x55mm REF: CTB23

FDA recall Class II

Ongoing

FDA lists this Class II recall for Kii Low Profile, Bladed Dual Pack , Qty. 1 5x55mm REF: CTB23. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1010-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 11, 2025
FDA report date
Jan 14, 2026
Quantity in commerce
450 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Kii Low Profile, Bladed Dual Pack , Qty. 1 5x55mm REF: CTB23

Affected model & UDI checker

Search the exact identifiers published for Z-1010-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CTB23Kii Low Profile, Bladed Dual Pack , Qty. 1 5x55mm REF: CTB23Check source noticeNot specified
Not specifiedKii Low Profile, Bladed Dual Pack , Qty. 1 5x55mm REF: CTB23Check source notice00607915126582
FDA code information

UDI: (01)00607915126582; Lot # 1546141

View the original FDA recall notice

Why was this product recalled?

There is the potential for extended length of the blade obturator after assembly into the cannula and seal housing.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did APPLIED MEDICAL RESOURCES CORP. take?

On December 11, 2025, Applied Medical issued a Urgent Medical Device Recall Notification to affected consignees via Email. Applied Medical ask consignees to take the following actions: 1. Check your inventory for recalled product. 2. Quarantine any product from the impacted lot. 3. Complete the attached Customer Recall Notification Confirmation Form. Please note that you must return the form even if you have no devices in inventory. 4. Please notify any facilities to which you distributed the affected product with a copy of this notification. Please also complete the Distributor Recall Notification Confirmation Report. 5. Return the Recall Notification Confirmation Form to Applied Medical by emailing to [email protected] 6. Return affected product and a copy of the Recall Notification Confirmation Form to Applied Medical

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology