Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1012-2026

MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L Recall

MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L

FDA recall Class I

Ongoing

FDA lists this Class I recall for MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1012-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Nov 26, 2025
FDA report date
Jan 28, 2026
Quantity in commerce
13245 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L

Affected model & UDI checker

Search the exact identifiers published for Z-1012-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MDR107002LMEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002LCheck source noticeNot specified
Not specifiedMEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002LCheck source notice40080196294320
FDA code information

UDI/DI 40080196294320, All lots

View the original FDA recall notice

Why was this product recalled?

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medline Industries, LP take?

Medline issued an URGENT MEDICAL DEVICE RECALL notice with an enclosed URGENT MEDICAL DEVICE FIELD CORRECTION NOTICE to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology