MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L Recall
MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L
FDA recall Class I
Ongoing
FDA lists this Class I recall for MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1015-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Nov 26, 2025
- FDA report date
- Jan 28, 2026
- Quantity in commerce
- 35694 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L
Affected model & UDI checker
Search the exact identifiers published for Z-1015-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MDR107003L | MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L | Check source notice | Not specified |
| Not specified | MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L | Check source notice | 40080196294344 |
FDA code information
UDI/DI 40080196294344, All lots
Why was this product recalled?
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Medline Industries, LP take?
Medline issued an URGENT MEDICAL DEVICE RECALL notice with an enclosed URGENT MEDICAL DEVICE FIELD CORRECTION NOTICE to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1015-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology