Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1040-2026

Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System. Recall

Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1040-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 4, 2025
FDA report date
Jan 21, 2026
Quantity in commerce
101 units
Distribution
See key details
FDA cause
Labeling Change Control

What is being recalled?

Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Part No. MI-76-0237; UDI: (01) 00869346000200 (11)230920 (21) AIRO-XXXX; Serial No. 2232206713, 2232206723, 2232206753, 2232206763, 2303430363, 2303430373, 2303430383, 2303430393, 2304034583, 2304034593, 2304034603, 2304034613, 2304034623, 2305514123, 2305514133, 2305514143, 2305514153, 2311019263, 2315101273, 2315101283, 2315101293, 2315101303, 2316030903, 2316030913, 2316703793, 2316703803, 2316703813, 2316703823, 2317415513, 2317415523, 2317415533, 2317415543, 2317415553, 2317905083, 2317905093, 2317905103, 2317905113, 2317905123, 2323513673, 2323513683, 2323513693, 2331419663, 2331419673, 2332117403, 2332117413, 2332117423, 2332117433, 2332117443, 2333404593, 2333404603, 2335506763, 2335506773, 2401909563, 2401909573, 2401909583, 2401909593, 2401909603, 2401909613, 2402911603, 2402911613, 2402911623, 2402911633, 2404302243, 2404302253, 2404302263, 2404302273, 2409620273, 2409620283, 2409620293, 2409620303, 2422207853, 2422207863, 2422207873, 2422207883, 2422207893, 2422207903, 2422207913, 2426303233, 2426303243, 2426303253, 2426303263, 2426303273, 2426303283, 2431210783, 2431210793, 2431210803, 2431210813, 2431210823, 2434504553, 2434504563, 2434504573, 2502103553, 2502103583, 2502103593, 2503121733, 2503121743, 2503121753, 2503616113, 2503616123, 2503616133, 2503616213.

View the original FDA recall notice

Why was this product recalled?

AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.

What actions did Mobius Imaging, LLC take?

An URGENT MEDICAL DEVICE CORRECTION notice dated 12/4/25 instructs consignees to ensure the following mitigations are in place while the correction is deployed: the hand position warnings as indicated in the IFU, train operators, the AIRO is accessible through a physical key switch and password protected log in screen, and the Backing Up System Warning label is affixed on to the gantry of affected devices. Stryker will correct devices during an upcoming service visit. Consignees are to inform users of devices of this recall by sharing the recall notice, immediately check stock areas and operating rooms to identify affected devices, and inform all users of the potential harms and their risk mitigations. Consignees with any questions are to contact their local sales office or Stryker Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.