Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1045-2025

Discovery XR656 HD. X-Ray imaging system. Recall

Discovery XR656 HD. X-Ray imaging system.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Discovery XR656 HD. X-Ray imaging system.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1045-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 28, 2025
FDA report date
Feb 12, 2025
Quantity in commerce
3149 (1697 US; 1452 OUS)
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

Discovery XR656 HD. X-Ray imaging system.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Definium Tempo (GTIN: 00195278118356) Definium Tempo Pro (GTIN: 00195278070265)

View the original FDA recall notice

Why was this product recalled?

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.

What actions did GE Medical Systems, LLC take?

Firm sent an "IMPORTANT ELECTRONIC PRODUCT RADIATION WARNING" email/letter to its consignees. The email/letter described the product, the problem and the actions to be taken. The consignees were instructed to do the following: You can continue to use the device as intended for clinical use, by following the instructions below: 1.If you perform an exposure and observe an on-screen message that the AEC termination limit has been reached, please acknowledge the message. 2.Take any relevant actions to adjust techniques or patient positioning for the next exposure to avoid reaching the termination limit for that exposure, if appropriate. Please ensure that all potential users in your facility are made aware of this notification and the recommended actions. Please retain this document for your records. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.