regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery. Recall
regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.
FDA recall Class I
Ongoing
FDA lists this Class I recall for regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1047-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Dec 26, 2024
- FDA report date
- Feb 12, 2025
- Quantity in commerce
- 414 kits
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.
Affected model & UDI checker
Search the exact identifiers published for Z-1047-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery. | Check source notice | 10194717119074 |
FDA code information
Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486.
Why was this product recalled?
There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did ROi CPS LLC take?
On 12/26/2024, the recalling firm issued a notice dated 12/26/2024 via email to their sole consignee notifying them of the recall of the convenience kits. The email also contained the notice issued by the manufacturer of the affected device component. The recalling firm's notice explained the reason for recall, the health risk involved, and the actions to be taken by the consignee, which included: (1) CHECK: Please review your inventory to determine if the manufacturer item numbers and lot numbers listed in the above table are in your inventory. (2) QUARANTINE: Please quarantine the kits affected by this notice and do not further distribute to end user facilities. (3) NOTIFY: Immediately notify any customers to whom you have distributed or forwarded product affected by this medical device correction notice. Provide a copy of this notice to all customers who have received impacted product and require a response. Customer responses will be requested to determine effectiveness of this recall. (4) END USERS: End users are instructed to discontinue use of the affected kits and quarantine. Kits are to be returned to the distribution location. (5) REPLY: Reply to this notice by completing the reply form attached and emailing to [email protected]. (6) CORRECTION: ROi CPS, LLC personnel will conduct correction activities at the distribution location to ensure removal of the recalled Mercury Medical item. The enclosed reply form is for the consignee to report they have read and understand the attached notice. The consignee is to indicate whether they have forwarded the notice to all customers who have been shipped the affected product, whether or not they have the consignee has the affected product, and that they have completed the "Quantity on Hand" portion of the reply form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1047-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.