FDA RECALL RECORD · Z-1066-2025

Infinity Central Station. Recall

Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Infinity Central Station.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1066-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 14, 2025
FDA report date
Feb 5, 2025
Quantity in commerce
1,013 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.

Affected model & UDI checker

Search the exact identifiers published for Z-1066-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MS26800Infinity Central Station.Check source noticeNot specified
MS34036Infinity Central Station.Check source noticeNot specified
Not specifiedInfinity Central Station.Check source notice04049098001878
Not specifiedInfinity Central Station.Check source notice04048675548027
FDA code information

Infinity CentralStation (ICS) Model Number: MS26800. UDI-DI Number: 04049098001878. Edifier Speaker Model Number: MS34036. UDI-DI Number: 04048675548027. Serial Numbers: 12946M00269, 12946M00314, 12946M00506, 12946M00519, 12946M00518, 12946M00553, 1OQ7WR-00103, 1OQ8CA-00249, 1OQ8CA-00254, 1OQ8CA-00229, 1OQ8CA-00221, 1OQ8CA-00282, 1OQ8CA-00314, 1OQ8CA-00193, 1OQ8CA-00200, 1OQ8CA-00194, 1OQ8CA-00181, 1OQ8CA-00178, 1OQ8CA-00248, 1OQ8CA-00125, 1OQ8CA-00071, 1OQ8CA-00032, 1OQ8CA-00171, 1OQ8CA-00164, 1OQ8CA-00149, 1OQ8CA-00069, 1OQ8CA-00057, 1OQ8CA-00153, 1OQ8CA-00150, 1OQ8CA-00148, 1OQ8CA-00165, 1OQ8CA-00155, 1OQ8CA-00097, 1OQ8CA-00113, 1OQ8CA-00106, 1OQ8CA-00075, 1OQ8CA-00052, 1OQ8CA-00067, 1OQ8CA-00128, 1OQ8CA-00132, 1OQ8CA-00268, 1OQ8CA-00198, 1OQ8CA-00255, 1OQ8CA-00151, 1OQ8CA-00342, 1OQ8CA-00391, 1OQ8CA-00321, 1OQ8CA-00402, 1OQ8CA-00395, 1OQ8CA-00297, 1OQ8CA-00240, 1OQ89M-00691, 1OQ89M-00690, 1OQ89M-00708, 1OQ89M-00693, 1OQ89M-00687, 1OQ89M-00707, 1OQ89M-00703, 1OQ89M-00679, 1OQ89M-00683, 1OQ89M-00692, 1OQ89M-00610, 1OQ89M-00613, 1OQ89M-00669, 1OQ89M-00635, 1OQ89M-00664, 1OQ89M-00697, P1OQ84F-00080, 1OQ89M-00684, 1OQ89M-00676, 1OQ89M-00695, 1OQ8CA-00086, 1OQ89M-00677, 1OQ8CA-00031, 1OQ8CA-00027, 1OQ8CA-00065, 1OQ8CA-00092, 1OQ8CA-00082, 1OQ89M-00602, 1OQ89M-00680, 1OQ89M-00675, 1OQ89M-00593, 1OQ89M-00686, 1OQ89M-00696, 1OQ89M-00681, 1OQ89M-00704, 1OQ89M-00674, 1OQ89M-00661, 1OQ89M-00663, P1OQ89M-00559, P1OQ89M-00540, 1OQ8CA-00134, 1OQ8CA-00095, 1OQ8CA-00043, 1OQ8CA-00074, 1OQ8CA-00762, 1OQ8CA-00688, 1OQ8CA-00670, 1OQ8CA-00637, 1OQ8CA-00633, 1OQ8E8-00037, 1OQ8E8-00053, 1OQ8CA-00773, 1OQ8CA-00683, 1OQ8CA-00500, 1OQ8CA-00476, 1OQ8CA-00449, 1OQ8CA-00420, 1OQ8CA-00460, 1OQ8CA-00463, 1OQ8CA-00507, 1OQ8CA-00350, 1OQ8CA-00442, 1OQ8CA-00433, 1OQ8CA-00447, 1OQ8CA-00601, 1OQ8CA-00604, 1OQ8CA-00695, 1OQ8CA-00513, 1OQ8CA-00575, 1OQ8CA-00577, 1OQ8CA-00591, 1OQ8CA-00197, 1OQ8CA-00516, 1OQ8CA-00541, 1OQ8CA-00542, 1OQ8CA-00568, 1OQ8CA-00499, 1OQ8CA-00498, P12051M00227, 1OQ8CA-00593, 12946M00532, 12946M00256, 12946M00278, 1OQ7WR-00203

View the original FDA recall notice

Why was this product recalled?

Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.

What actions did Draeger Medical Systems, Inc. take?

Draeger notified consignees on about 01/14/2025 via Urgent - Medical Device Correction letter. Consignees were instructed to inspect the device to identify if the speaker is configured with the ICS in USB mode or Analog mode. For the speaker set to Analog mode there is an analog cable (aux cable) connected to the backside of the speaker and a light on frontside of the speaker display. If the light is green, there is no issue of speaker audio annunciation. If the light is red, the issue of speaker audio annunciation is present. If an issue with alarm annunciation occurs while operating, follow the steps included in the letter. Ensure all affected users or customers, if further distributed, are notified of the recall notification. They were instructed to complete and return the provided reply card.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.