FDA RECALL RECORD · Z-1072-2025

SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS). Recall

SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).

FDA recall Class II

Ongoing

FDA lists this Class II recall for SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1072-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 12, 2024
FDA report date
Feb 12, 2025
Quantity in commerce
81,606
Distribution
See key details
FDA cause
Device Design

What is being recalled?

SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).

Affected model & UDI checker

Search the exact identifiers published for Z-1072-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1000SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).Check source noticeNot specified
1000-DSenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).Check source noticeNot specified
FDA code information

REF/UDI-DI: 1000/05425025750405, 1000-D/05425025750528 All Serial Numbers less than 500,000

View the original FDA recall notice

Why was this product recalled?

Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did LivaNova USA, Inc. take?

On 12/12/2024, correction notices were mailed to customers asking them to do the following: 1. Monitor patient frequently and perform an on-demand System Diagnostic test at each visit per labeling (typical patient visit cadence is every 3-12 months). See VNS Therapy Physician's Manual, found in the Manuals Section https://www.livanova.com/epilepsy-vnstherapy/en-us/hcp/product-manuals or https://www.livanova.com/depression/en-us/healthcare-provider-resources. 2. Counsel prospective and current patients to ensure that they are aware of and understand the risks. You may elect to provide the patient letter. 3. Contact Customer Quality at 1-866-882-8804 or e-mail at [email protected] if communication with the generator using the Programming System indicates that: i. Diagnostic tests cannot be performed (Error Code 254 - Test Interrupted error message) ii. System Diagnostic test displays a LOW lead impedance result. 4. Encourage patients to: i. Notify you if there is a change in perceived clinical symptoms (e.g. increase in seizures or depressive symptoms, etc.). ii. Notify you if they no longer perceive any form of stimulation. 5. Ensure that this notice is communicated to all personnel within your organization who need to be aware. 6. Complete and return the response form via email to [email protected] Firm is updating the VNS Therapy Instructions for Use (IFU) to reflect the patient management recommendations relevant to this issue. For questions contact the firm's Customer Quality at 1-866-882-8804 or e-mail [email protected] On 6/16/2025, updated recall notices were delivered in person notifying customers that devices with a design change were available so affected generators remaining in inventory should be replaced. New devices with the design change have serial numbers greater than 500,000. There are no new actions for devices that have already been implanted.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology