MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO) Recall
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
FDA recall Class II
Ongoing
FDA lists this Class II recall for MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1078-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 10, 2025
- FDA report date
- Jan 21, 2026
- Quantity in commerce
- 1
- Distribution
- See key details
- FDA cause
- Software change control
What is being recalled?
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Affected model & UDI checker
Search the exact identifiers published for Z-1078-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO) | Check source notice | 07340201500026 |
| Not specified | MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO) | Check source notice | 07340201500071 |
FDA code information
MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071
Why was this product recalled?
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Elekta, Inc. take?
On 12/10/2025 correction notices were emailed to consignees who were asked to do the following: Radiation therapists must always check the meterset values displayed on the MOSAIQ console with the value displayed on the treatment unit console. If a mismatch occurs, follow Caution 9.6 in the MOSAIQ Instructions for Use. Complete and return the acknowledgement form. Firm is working on the correction and will release a Field Safety Modification with the details as soon as possible. Elekta Regional Contact Details: Region America: [email protected] Tel: +1 855 6935358 https://www.elekta.com Region Asia Pacific: [email protected] Tel: + 852 2891 2208 https://www.elekta.com Japan [email protected] Tel: + 81 3 6722 3800 Fax: +81 3 6436 4231 https://www.elekta.com Region China: [email protected] Tel : + 86 10 8012 5012 https://www.elekta.com Region Europe: [email protected] Tel : + 46 8 587 254 00 https://www.elekta.com Region Turkey, India & Middle East [email protected] Turkey Tel: +90 216 444 6374 India Tel: +1-800-103-7454 Middle East Tel: +00 800 4000 5000 https://www.elekta.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1078-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology