Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG Recall
Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG
FDA recall Class II
Ongoing
FDA lists this Class II recall for Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1100-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 19, 2025
- FDA report date
- Jan 21, 2026
- Quantity in commerce
- 64290 total
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG
Affected model & UDI checker
Search the exact identifiers published for Z-1100-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG | Check source notice | 30351688073062 |
| Not specified | Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG | Check source notice | 30351688006176 |
| Not specified | Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG | Check source notice | 30351688029540 |
| Not specified | Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG | Check source notice | 30351688039150 |
| Not specified | Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG | Check source notice | 30351688081173 |
FDA code information
1. 4044EG-1 UDI-DI 30351688073062 Lots 6141992 6156228 2. 4080P-2 UDI-DI 30351688006176 Lots 6141999 6142000 6147906 3. 4084P-1 UDI-DI 30351688029540 Lots 6147907 (corrected 1/22/26, removed "-LSM" suffix) 4. 4080PE UDI-DI 30351688039150 Lot 6131839 5. 4080PG UDI-DI 30351688081173 Lots 6131838 6142001 6142002
Why was this product recalled?
Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.
What actions did Icu Medical, Inc. take?
On December 19. 2025, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were informed that affected lots may have a crack present on the syringe collar. Customers were instructed to discontinue use of affected lots, and to destroy products following their institution's process for destruction. If destruction is not immediately possible, affected product should be quarantined until disposal.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1100-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology