Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1102-2025

LIFEPAK 35 ECG cable REF 11111-000041 Recall

LIFEPAK 35 ECG cable REF 11111-000041

FDA recall Class II

Ongoing

FDA lists this Class II recall for LIFEPAK 35 ECG cable REF 11111-000041. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1102-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 21, 2025
FDA report date
Feb 12, 2025
Quantity in commerce
245 units
Distribution
See key details
FDA cause
Labeling Change Control

What is being recalled?

LIFEPAK 35 ECG cable REF 11111-000041

Affected model & UDI checker

Search the exact identifiers published for Z-1102-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
11111-000041LIFEPAK 35 ECG cable REF 11111-000041Check source noticeNot specified
Not specifiedLIFEPAK 35 ECG cable REF 11111-000041Check source notice00883873834596
FDA code information

Catalog Number: 11111-000041 UDI-DI code: 00883873834596 Lot/Serial Number: 0224

View the original FDA recall notice

Why was this product recalled?

Product shipped with incorrect IFU. IFU does not include proper cleaning/disinfection instructions

What actions did Physio-Control, Inc. take?

On 01/21/2025, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that Stryker discovered that an incorrect version of the Instructions for Use (IFU) for the 3-wire extended precordial ECG cable for the LIFEPAK 35 were distributed to customers. This version did not include information regarding disinfection of the 3-wire cable. Customer are instructed to: 1. Immediately check internal inventory to locate the product listed on the attached business reply form. 2. Remove and quarantine any affected product from use. 3. To confirm receipt of this Medical Device Correction Notice and understanding of the provided information, email the enclosed Business Reply Form (BRF) by January 31st, 2025 date, to [email protected]. 4. Upon receipt of the completed business reply form, Stryker will contact customers to arrange for the return and replacement of their product(s). 5. Maintain awareness of this communication internally until all required actions have been completed within their facility. Ensure this letter is kept with the affected device until replacement. 6. Inform Stryker if any of the subject devices have been distributed to other organizations by contacting Stryker at: a. Device Tracking: [email protected] b. Website: https://www.strykeremergencycare.com/service-support-overview/device-registration/ If you have any questions or concerns, contact Customer Service +1 800 442 1142 or by email [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology