IntelliVue Patient Monitor MX450. Recall
IntelliVue Patient Monitor MX450. Model Number: 866062
FDA recall Class II
Ongoing
FDA lists this Class II recall for IntelliVue Patient Monitor MX450.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1108-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 15, 2025
- FDA report date
- Feb 12, 2025
- Quantity in commerce
- 98 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
IntelliVue Patient Monitor MX450. Model Number: 866062
Affected model & UDI checker
Search the exact identifiers published for Z-1108-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 866062 | IntelliVue Patient Monitor MX450. | Check source notice | Not specified |
| Not specified | IntelliVue Patient Monitor MX450. | Check source notice | 00884838038769 |
| Not specified | IntelliVue Patient Monitor MX450. | Check source notice | 884838045859 |
FDA code information
Model Number: 866062. UDI-DI: (01)00884838038769(21); Serial Numbers: DE35169761, DE671V3903, DE671V3885, DE671V7898, DE7770DC1L. UDI-DI: 884838045859; Serial Numbers: DE32303399, DE32302479, DE32302480, DE32306956, DE35131294, DE35114613, DE35114612, DE671T9641, DE7770C86L, DE7770C86K, DE7770C86J, DE7770C86H, DE7770C86G, DE35120431, DE671K3731, DE671K3729, DE671K3706, DE671K3796, DE671K3816, DE671K3737, DE671K3773, DE671L0436, DE671L0462, DE671L0406, DE671L5214, DE671L5207, DE671L5206, DE671L5181, DE671L5169, DE671M3993, DE671M4048, DE671M4036, DE671M4037, DE671M4053, DE671M7009, DE671M7012, DE671M9968, DE671M9972, DE671M9963, DE671N0009, DE671N0010, DE671N0092, DE671N0008, DE671N0001, DE671N0002, DE671M9985, DE671M9999, DE671N0000, DE671P0411, DE671P1094, DE671P2457, DE671P2537, DE32306016, DE32306017, DE32306018, DE7580B4PC, DE7580B4PB, DE7580B4PA, DE7770C6ZP, DE7770C74Y, DE7770C74Z, DE7770C750, DE7770C74W, DE7770C74V, DE7770C78E, DE7770C78D, DE7770D6LH, DE7770D6LG, DE7770D6LF, DE7770D6LE, DE35142704, DE35142703, DE35142705, DE35142702, DE35142706, DE35182894, DE32301491, DE32301476, DE32301477, DE32301478, DE32301479, DE32301480, DE32301481, DE32301482, DE32301483, DE32301484, DE32301485, DE32301486, DE32301487, DE32301488, DE32301489, DE32301490, DE32301506.
Why was this product recalled?
Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Philips North America Llc take?
Philips began notifying consignees on about 01/15/2025 via certified letters titled URGENT Field Safety Notice and URGENT Medical Device Correction letters. Consignees were instructed to ensure that there is no interruption of the cable connection between the Hamilton ventilator and the EC10 Module or IntelliBridge I/O board before use, provide the notification to all affected personnel and customers if the devices have been further distributed, and to complete and return the response form provided. A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will upgrade your ED245 Driver from revision A.0 to revision A.1. Distributors were instructed to return the Response Form attached your contact information, send the attached URGENT Field Safety Notice to each customer to whom you have distributed any affected device as soon as possible and no later than 5 days, together with the Reply Card, perform a good faith effort to get the Response Form by following up with the customer with a minimum of three attempts, and if possible, using multiple contact methods, inform Philips about the responses received, and implement the Technical solution established by Philips as soon as available within the timeframe communicated by Philips and confirm to Philips the implementation in all affected products. In case a product is not corrected, inform Philips on the reason why the technical solution cannot be implemented.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1108-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology