FDA RECALL RECORD · Z-1110-2025

IntelliVue Patient Monitor MX550. Recall

IntelliVue Patient Monitor MX550. Model Number: 866066

FDA recall Class II

Ongoing

FDA lists this Class II recall for IntelliVue Patient Monitor MX550.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1110-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 15, 2025
FDA report date
Feb 12, 2025
Quantity in commerce
74 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

IntelliVue Patient Monitor MX550. Model Number: 866066

Affected model & UDI checker

Search the exact identifiers published for Z-1110-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
866066IntelliVue Patient Monitor MX550.Check source noticeNot specified
Not specifiedIntelliVue Patient Monitor MX550.Check source notice00884838038783
FDA code information

Model Number: 866066. UDI-DI: (01)00884838038783(21); Serial Numbers: DE671E3020, DE671E2989, DE671E2988, DE671E3086, DE671E3085, DE671E3084, DE671E3097, DE671E3096, DE671E3095, DE671E3094, DE671E3093, DE671E3253, DE671E3246, DE671E3245, DE671E3244, DE671E3241, DE671E3240, DE671E3234, DE671E3235, DE671E3238, DE671E3239, DE671E3021, DE671E3236, DE671E3865, DE671E3863, DE671E3862, DE671E3861, DE671E3860, DE671E3859, DE671E3858, DE671E3857, DE671E3856, DE671E3855, DE671E3853, DE671E3852, DE671E3867, DE671E3871, DE671E3730, DE671E3729, DE671E3866, DE35186967, DE35186949, DE35186948, DE35186947, DE35186944, DE35186943, DE35186942, DE35186941, DE35186940, DE35186939, DE35186938, DE35186982, DE35186945, DE35199396, DE35199398, DE35199428, DE671E2030, DE671H1334, DE32302561, DE32302562, DE32301002, DE32301003, DE32301004, DE35114609, DE35114608, DE758Y7635, DE758Y7616, DE758Y7619, DE758Y7615, DE758Y7614, DE7580AWK6, DE7580AWK7, DE7580AWKB, DE7580AWK4.

View the original FDA recall notice

Why was this product recalled?

Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America Llc take?

Philips began notifying consignees on about 01/15/2025 via certified letters titled URGENT Field Safety Notice and URGENT Medical Device Correction letters. Consignees were instructed to ensure that there is no interruption of the cable connection between the Hamilton ventilator and the EC10 Module or IntelliBridge I/O board before use, provide the notification to all affected personnel and customers if the devices have been further distributed, and to complete and return the response form provided. A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will upgrade your ED245 Driver from revision A.0 to revision A.1. Distributors were instructed to return the Response Form attached your contact information, send the attached URGENT Field Safety Notice to each customer to whom you have distributed any affected device as soon as possible and no later than 5 days, together with the Reply Card, perform a good faith effort to get the Response Form by following up with the customer with a minimum of three attempts, and if possible, using multiple contact methods, inform Philips about the responses received, and implement the Technical solution established by Philips as soon as available within the timeframe communicated by Philips and confirm to Philips the implementation in all affected products. In case a product is not corrected, inform Philips on the reason why the technical solution cannot be implemented.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.